NZS/ISO/IEC 17025: 2005 General requirements for the competence of testing and calibration laboratories. This page details only those capabilities covered under our IANZ accredition. Please contact us for full details of our capabilities. Nominal calibration temperature: (23 ± 5) °C. All best measurement capabilities (BMCs) are based on a Confidence Interval of at least 95%. Unless stated elsewhere in this schedule, calibrations are performed at the premises of the accredited laboratory. 5.88 Calibrators for Instrumentation(a) DC Voltage
| 0 mV to 120 mV | ± (22 ppm + 1.3 µV) | | 0 V to 1.2 V | ± (14 ppm + 2 µV) | | 0 V to 12 V | ± (13 ppm + 3 µV) | | 0 V to 120 V | ± (16 ppm + 40 µV) | | 0 V to 1050 V | ± (10.5 ppm + 100 µV) |
(b) AC Voltage
| 1.2 mV to 12 mV | 1 Hz to <40 Hz | ± (0.11% + 3 µV) | | | 40 Hz to <1 kHz | ± (0.07% + 1.1 µV) | | | 1 kHz to <20 kHz | ± (0.22% + 1.1 µV) | | | 20 kHz to 300 kHz | ± (0.6% + 1.1 µV)* | | 12 mV to 120 mV | 1 Hz to <40 Hz | ± (0.077% + 4 µV) | | | 40 Hz to <1 kHz | ± (0.024% + 2 µV) | | | 1 kHz to <20 kHz | ± (0.048% + 2 µV) | | | 20 kHz to 2 MHz | ± (0.08% + 2 µV)* | | 120 mV to 1.2 V | 1 Hz to <40 Hz | ± (0.068% + 40 µV) | | | 40 Hz to <1 kHz | ± (0.011% + 20 µV) | | | 1 kHz to <20 kHz | ± (0.017% + 20 µV) | | | 20 kHz to 2 MHz | ± (0.04% + 20 µV)* | | 1.2 V to 12 V | 1 Hz to <40 Hz | ± (0.068% + 400 µV) | | | 40 Hz to <1 kHz | ± (0.012% + 200 µV) | | | 1 kHz to <20 kHz | ± (0.018% + 200 µV) | | | 20 kHz to 2 MHz | ± (0.04% + 200 µV)* | | 12 V to 120 V | 1 Hz to <40 Hz | ± (0.08% + 4 mV) | | | 40 Hz to <1 kHz | ± (0.03% + 2 mV) | | | 1 kHz to <20 kHz | ± (0.03% + 2 mV) | | | 20 kHz to 1 MHz | ± (0.039% + 2 mV)* | | 120 V to 700 V | 1 Hz to <40 Hz | ± (0.05% + 40 mV) | | | 40 Hz to <1 kHz | ± (0.05% + 20 mV) | | | 1 kHz to <20 kHz | ± (0.07% + 20 mV) | | | 20 kHz to 100 kHz | ± (0.13% + 20 mV)* |
This is an abbreviated table. BMCs marked * are the best available over several discrete frequency ranges combined (values given are taken from the lower end of the frequency range).
(c) DC Current
| 0 nA to 120 nA | ± (560 ppm + 40 nA) | | 0 µA to 1.2 µA | ± (84 ppm + 40 nA) | | 0 µA to 12 µA | ± (36 ppm + 100 nA) | | 0 µA to 120 µA | ± (35 ppm + 800 nA) | | 0 mA to 1.2 mA | ± (37 ppm + 0.005 µA) | | 0 mA to 12 mA | ± (39 ppm + 0.05 µA) | | 0 mA to 120 mA | ± (49 ppm + 0.5 µA) | | 0 A to 1.05 A | ± (120 ppm + 10 µA) |
(d) AC Current
| 2.4 µA to 120 µA | 10 Hz to <20 Hz | ± (0.5% + 0.03 µA) | | | 20 Hz to <45 Hz | ± (0.18% + 0.03 µA) | | | 45 Hz to <100 Hz | ± (0.08% + 0.03 µA) | | | 100 Hz to 1 kHz | ± (0.08% + 0.03 µA)* | | 0.06 mA to 1.2 mA | 10 Hz to <20 Hz | ± (0.5% + 0.2 µA) | | | 20 Hz to <45 Hz | ± (0.17% + 0.2 µA) | | | 45 Hz to <100 Hz | ± (0.07% + 0.2 µA) | | | 100 Hz to 1 kHz | ± (0.04% + 0.2 µA) | | | >1 kHz to <5 kHz | ± (0.09% + 0.2 µA) | | | 5 kHz to <50 kHz | ± (0.28% + 0.2 µA)* | | | 50 kHz to 100 kHz | ± (2.52% + 1.5 µA) | | 0.6 mA to 12 mA | 10 Hz to <20 Hz | ± (0.5% + 2 µA) | | | 20 Hz to <45 Hz | ± (0.17% + 2 µA) | | | 45 Hz to <100 Hz | ± (0.07% + 2 µA) | | | 100 Hz to 1 kHz | ± (0.04% + 2 µA) | | | >1 kHz to <5 kHz | ± (0.09% + 2 µA) | | | 5 kHz to <50 kHz | ± (0.28% + 2 µA)* | | | 50 kHz to 100 kHz | ± (2.5% + 15 µA) | | 6mA to 120mA | 10 Hz to <20 Hz | ± (0.5% + 20 µA) | | | 20 Hz to <45 Hz | ± (0.17% + 20 µA) | | | 45 Hz to <100 Hz | ± (0.07% + 20 µA) | | | 100 Hz to 1 kHz | ± (0.04% + 20 µA) | | | >1 kHz to <5 kHz | ± (0.1% + 20 µA) | | | 5k Hz to <50 kHz | ± (0.3% + 20 µA) | | | 50 kHz to 100 kHz | ± (2.8% + 150 µA) | | 52.5mA to 1.05A | 10 Hz to <20 Hz | ± (0.9% + 200 µA) | | | 20 Hz to <45 Hz | ± (0.34% + 200 µA) | | | 45 Hz to <100 Hz | ± (0.17% + 200 µA) | | | 100 Hz to 1 kHz | ± (0.2% + 200 µA) | | | >1 kHz to <5 kHz | ± (0.3% + 200 µA) | | | 5 kHz to <20 kHz | ± (1.1% + 200 µA) | | | 20 kHz to 50 kHz | ± (4% + 400 µA) |
(e) Resistance
| 0 Ohm to 12 Ohm | ± (39 ppm + 0.06 mOhm) | | 0 Ohm to 120 Ohm | ± (23 ppm + 0.5 mOhm) | | 0 Ohm to 1.2 kOhm | ± (19 ppm + 0.5 mOhm) | | 0 Ohm to 12 kOhm | ± (18 ppm + 5 Ohm) | | 0 Ohm to 120 kOhm | ± (21 ppm + 50 Ohm) | | 0 Ohm to 1.2 MOhm | ± (27 ppm + 2 kOhm ) | | 0 Ohm to 12 MOhm | ± (67 ppm + 100 kOhm) | | 0 Ohm to 120 MOhm | ± (520 ppm + 1 MOhm) | 0 Ohm to 1.2 GOhm | ± (5000 ppm + 10 MOhm) |
TOP 5.89 Indicating Instruments and Recording Instruments(a) DC Voltmeters
| 0 mV to <330 mV | ± (60 ppm + 3 µV) | | 330 mV to <3 V | ± (50 ppm + 5 µV) | | 3 V to <30 V | ± (50 ppm + 50 µV) | | 30 V to <100 V | ± (55 ppm + 500 µV) | | 100 V to 1020 V | ± (55 ppm + 1500 µV) |
(b) AC Voltmeters
| 1 mV to <33 mV | 0.01 Hz to <10 Hz | ± (5% + 165 µV) | | | 10 Hz to <20 Hz | ± (0.84% + 20 µV) | | | 20 Hz to <45 Hz | ± (0.35% + 20 µV) | | | 45 Hz to <10 kHz | ± (0.15% + 20 µV) | | | 10 kHz to <20 kHz | ± (0.2% + 20 µV) | | | 20 kHz to <50 kHz | ± (0.25% + 20 µV) | | | 50 kHz to <100 kHz | ± (0.35% + 33 µV) | | | 100 kHz to 500 kHz | ± (1.4% + 60 µV) | | 33 mV to <330 mV | 0.01 Hz to <10 Hz | ± (5% + 1650 µV) | | | 10 Hz to <20 Hz | ± (0.75% + 50 µV) | | | 20 Hz to <45 Hz | ± (0.25% + 50 µV) | | | 45 Hz to <10 kHz | ± (0.05% + 20 µV) | | | 10 kHz to <20 kHz | ± (0.1% + 20 µV) | | | 20 kHz to <50 kHz | ± (0.16% + 40 µV) | | | 50 kHz to <100 kHz | ± (0.24% + 170 µV) | | | 100 kHz to 500 kHz | ± (7.7% + 330 µV) | | 330 mV to <3.3 V | 0.01 Hz to <10 Hz | ± (5% + 16.5 mV) | | | 10 Hz to <20 Hz | ± (0.16% + 250 µV) | | | 20 Hz to <45 Hz | ± (0.15% + 250 µV) | | | 45 Hz to <10 kHz | ± (0.03% + 60 µV) | | | 10 kHz to <20 kHz | ± (0.08% + 60 µV) | | | 20 kHz to <50 kHz | ± (0.14% + 300 µV) | | | 50 kHz to <100 kHz | ± (0.24% + 1700 µV) | | | 100 kHz to 500 kHz | ± (7.7% + 3300 µV) | | 3.3 V to 33 V | 0.01 Hz to <10 Hz | ± (5% + 165 mV) | | | 10 Hz to <20 Hz | ± (0.17% + 2500 µV) | | | 20 Hz to <45 Hz | ± (0.15% + 2500 µV) | | | 45 Hz to <10 kHz | ± (0.04% + 600 µV) | | | 10 kHz to <20 kHz | ± (0.08% + 2600 µV) | | | 20 kHz to <50 kHz | ± (0.19% + 5000 µV) | | | 50 kHz to 100 kHz | ± (0.24% + 17 mV) | | 33 V to <330 V | 45 Hz to <1 kHz | ± (0.05% + 6600 µV) | | | 1 kHz to <10 kHz | ± (0.08% + 15 mV) | | | 10 kHz to 20 kHz | ± (0.09% + 33 mV) | | 330 V to 1020 V | 45 kHz to <1 kHz | ± (0.05% + 80 mV) | | | 1 kHz to <5 kHz | ± (0.2% + 100 mV) | | | 5 kHz to 10 kHz | ± (0.2% + 500 mV) |
(c) DC Ammeters
| 0 mA to <3.3 mA | ± (0.014% + 0.05 µA) | | 3.3 mA to <33 mA | ± (0.02% + 0.25 µA) | | 33 mA to <330 mA | ± (0.03% + 3.3 µA) | | 330 mA to <2.2 A | ± (0.03% + 44 µA) | | 2.2 A to 11 A | ± (0.06% + 330 µA) |
(d) AC Ammeters
| 0.029mA to 0.33mA | 0.01 Hz to <10 Hz | ± (5% + 1.7 µA) | | | 10 Hz to <20 Hz | ± (0.58% + 0.15 µA) | | | 20 Hz to <45 Hz | ± (0.21% + 0.15 µA) | | | 45 Hz to <65 Hz | ± (0.21% + 0.25 µA) | | | 65 Hz to <1 kHz | ± (0.13% + 0.25 µA) | | | 1 kHz to <5 kHz | ± (0.4% + 0.15 µA) | | | 5 kHz to 10 kHz | ± (1.4% + 0.15 µA) | | 0.33 mA to 3.3 mA | 0.01 Hz to <10 Hz | ± (5% + 17 µA) | | | 10 Hz to <20 Hz | ± (0.6% + 0.3 µA) | | | 20 Hz to <45 Hz | ± (0.1% + 0.3 µA) | | | 45 Hz to <65 Hz | ± (0.3% + 0.3 µA) | | | 65 Hz to <1 kHz | ± (0.1% + 0.3 µA) | | | 1 kHz to <5 kHz | ± (0.2% + 0.3 µA) | | | 5 kHz to 10 kHz | ± (0.7% + 0.3 µA) | | 3.3 mA to 33 mA | 0.01 Hz to <10 Hz | ± (5% + 170 µA) | | | 10 Hz to <20 Hz | ± (0.6% + 3 µA) | | | 20 Hz to <45 Hz | ± (0.1% + 3 µA) | | | 45 Hz to <65 Hz | ± (0.25% + 3 µA) | | | 65 Hz to <1 kHz | ± (0.09% + 3 µA) | | | 1 kHz to <5 kHz | ± (0.2% + 3 µA) | | | 5 kHz to 10 kHz | ± (0.7% + 3 µA) | | 33 mA to 330 mA | 0.01 Hz to <10 Hz | ± (5% + 1700 µA) | | | 10 Hz to <20 Hz | ± (0.4% + 30 µA) | | | 20 Hz to <45 Hz | ± (0.1% + 30 µA) | | | 45 Hz to <65 Hz | ± (0.27% + 30 µA) | | | 65 Hz to <1 kHz | ± (0.09% + 30 µA) | | | 1 kHz to <5 kHz | ± (0.3% + 30 µA)( | | | 5 kHz to 10 kHz | ± (1.3% + 30 µA) | | 0.33 mA to 2.2 A | 10 Hz to <65 Hz | ± (0.2% + 300 µA) | | | 65 Hz to <1 kHz | ± (0.1% + 300 µA) | | | 1 kHz to 5 kHz | ± (0.75% + 300 µA) | | 2.2 A to 11 A | 45 Hz to <65 Hz | ± (0.12% + 2000 µA) | | | 65 Hz to <500 Hz | ± (0.1% + 2000 µA) | | | 500 Hz to 1 kHz | ± (0.33% + 2000 µA) |
(e) WattmetersBest Measurement Capability for power (W and VA) is calculated as the root sum of squares of the BMCs for the appropriate voltage and current values (and power factor, if applicable). Voltage limitations: 33 mVac to 1000 Vac, or 0 V to 1000 Vdc. Current limitations: 3.3 mAac to 11 Aac, or 0 A to 11 Adc. Auxiliary voltage limitations: 10 mVac to 3.3 Vac, or 0 V to 3.3 Vdc. The range of direct output is from 108.9 µW through to 11 kW. Power can be simulated over greater ranges for equipment incorporating current clamps or current transformers. (Maximum output available for the voltage input is 1000 V, and maximum output available for the "current" input is 11 A or 3.3 V).
(f) VarmetersBest Measurement Capability for reactive power (VAr) is calculated as the root sum of squares of the BMCs for the appropriate voltage, current, and power factor values. Voltage and current limitations apply as for Wattmeters above.
(g) Phase Angle Indicators
| 0° to 179.98° | 10 Hz to <65 Hz | ± 0.15° | | | 65 Hz to <500 Hz | ± 0.9° | | | 500 Hz to <1 kHz | ± 2° | | | 1 kHz to <5 kHz | ± 6° | | | 5 kHz to 10 kHz | ± 10° |
Voltage and current limitations apply as for Wattmeters above.
(h) Power Factor MetersBest Measurement Capability for power factor may be calculated from the BMC given above for Phase angle indicators. [For example: BMC for power factor of 0.500 at 50Hz is ACOS (0.500) ± 0.15 which is 0.500 (+0.002266, -0.002269) or 0.500 ± 0.003]. Voltage and current limitations apply as for Wattmeters above.
(i) Ohmmeters
| 0 Ohm to <11 Ohm | ± (0.012% + 0.006 Ohm) | | 11 Ohm to <33 Ohm | ± (0.012% + 0.01 Ohm) | | 33 Ohm to <110 Ohm | ± (0.009% + 0.01 Ohm) | | 110 Ohm to <330 Ohm | ± (0.009% + 0.01 Ohm) | | 330 Ohm to <1.1 kOhm | ± (0.009% + 0.06 Ohm) | | 1.1 kOhm to <3.3 kOhm | ± (0.009% + 0.06 Ohm) | | 3.3 kOhm to <11 kOhm | ± (0.009% + 0.6 Ohm) | | 11 kOhm to <33 kOhm | ± (0.009% + 0.6 Ohm) | | 33 kOhm to <110 kOhm | ± (0.011% + 6 Ohm) | | 110 kOhm to <330 kOhm | ± (0.012% + 6 Ohm) | | 330 kOhm to <1.1 MOhm | ± (0.015% + 55 Ohm) | | 1.1 MOhm to <3.3 MOhm | ± (0.015% + 55 Ohm) | | 3.3 MOhm to <11 MOhm | ± (0.06% + 550 Ohm) | | 11 MOhm to <33 MOhm | ± (0.1% + 550 Ohm) | | 33 MOhm to <110 MOhm | ± (0.5% + 5.5 kOhm) | | 110 MOhm to 330 MOhm | ± (0.7% + 17 kOhm) |
(j) Capacitance Meters
| 0.33 nF to <11 nF | ± (0.5% + 0.01 nF) | | 11 nF to <110 nF | ± (0.25% + 0.1 nF) | | 110 nF to <330 nF | ± (0.25% + 0.3 nF) | | 330 nF to <1.1 µF | ± (0.25% + 1 nF) | | 1.1 µF to <3.3 µF | ± (0.35% + 3 nF) | | 3.3 µF to <11 µF | ± (0.35% + 10 nF) | | 11 µF to <33 µF | ± (0.4% + 30 nF) | | 33 µF to <110 µF | ± (0.5% + 100 nF) | | 110 µF to <330 µF | ± (0.7% + 300 nF) | | 330 µF to 1.1 mF | ± (1% + 300 nF) |
TOP (n) Digital Storage RecordersThis section gives BMCs for oscilloscope calibrations using the Wavetek 9500 calibrator.
| DC Voltage into 1 MOhm | | | -222.4 V to -888 µV | ± (0.025% + 25 µV) | | | | | 0 V | ± 15 µV | | | | | +888 µV to +222.4 V | ± (0.025% + 25 µV) | | DC Voltage into 50 Ohm | | | -5.56 V to -888 µV | ± (0.025% + 25 µV) | | | | | 0 V | ± 15 µV | | | | | +888 µV to +5.56 V | ± (0.025% + 25 µV) | | Square Voltage into 1 MOhm | (negative, positive or symmetric) | | 35.52 µVpp to 999.99 µVpp | ± (1% + 10 µV) | | | | | 1 mVpp to 21 mVpp | ± (0.1% + 15 µV) | | | | | 21.001 mVpp to 556 mVpp | ± (0.1% + 1 µV) | | | | | 556.01 mVpp to 210 Vpp | ± (0.05% + 1 µV) | | | | | 0V | ± 15 µV | | | *Output Frequency | | 10 Hz to 100 kHz | ± 0.25 ppm | | Square Voltage into 50 Ohm | (negative, positive, or symmetric) | | 35.52 µVpp to 999.99 µVpp | ± (1% + 10 µV)
| | | | | 1 mVpp to 21 mVpp | ± (0.1% + 15 µV) | | | | | 21.001 mVpp to 556 mVpp | ± (0.1% + 1 µV) | | | | | 556.01 mVpp to 5.56 Vpp | ± (0.05% + 1 µV) | | | | | 0V | ± 15µV | | | *Output Frequency | | 10 Hz to 100 kHz | ± 0.25 ppm | | Levelled Sine | *Reference level | 50 kHz to 10 MHz | 4.4 mVpp to 5.56 Vpp | ± 1.5% | | | *Flatness relative to Reference level | 100 mHz to 100 MHz | 4.4 mVpp to 5.56 Vpp | ± 1.5% | | | | 100 MHz to 550 MHz | 4.4 mVpp to 5.56 Vpp | ± 3% | | | | 550 MHz to 1.1 GHz | 4.4 mVpp to 3.336 Vpp | ± 4% | Low Edge into 50 Ohm and 1 MOhm | *Output Voltage | | 4.4 mVpp to 3.1 Vpp | ± 3% | | | *Rise/Fall Time (10% to 90%) | | 436 ps | ± 38ps | | | *Output Period | | 500 ns to 100 ms | ± 0.25 ppm | | High Edge into 50 Ohm | *Output Voltage | | 888 mVpp to 5.56 Vpp | ± 3% | | | *Rise/Fall Time (10% to 90%) | | 100 ns | Nominal | | | *Output Period | | 10 µs to 100 ms | ± 0.25 ppm | | High Edge into 1 MOhm | *Output Voltage | | 888 mVpp to 100 Vpp | ± 3% | | | | | 100.01 Vpp to 210 Vpp | ± 3% | | Rise/Fall Time (10% to 90%) | 888 mVpp to 100 Vpp
| | 150 ns | Nominal | | | 100.0 Vpp to 210 Vpp | | 200 ns | Nominal | | | *Output Period | | 10 µs to 100 ms | ± 0.25 ppm | | Fast Edge into 50 Ohm | *Output Voltage | | 4.4 mVpp to 3.1 Vpp | ± 3% | | | *Rise/Fall Time (10% to 90%) | | 149 ps | ± 29 ps | | | *Output Period | | 500 ns to 100 ms | ± 0.25 ppm | | Time Markers into 50 Ohm and 1 MOhm | *Square (nom.peak to peak) | 100 mV,250 mV,500 mV,1 V | 9.0091 ns to 55 s (Rise/Fall Time 1 ns typical) | ± 0.25 ppm | | | *Sine (nom.peak to peak) | 100 mV,250 mV,500 mV | 450.5 ps to 909.09 ps | ± 0.25 ppm | | | | 100 mV,250 mV,500 mV,1 V | 909.1 ps to 9.009 ns | ± 0.25 ppm | | | *Pulse (nom.peak to peak) | 100 mV,250 mV,500 mV,1 V | 900.91 ns to 55 s (Rise/Fall Time 1 ns typical) | ± 0.25 ppm | | | *Triangle(nom.peak to peak) | 100 mV,250 mV,500 mV,1 V | 900.91 ns to 55 s (Rise/Fall Time 2.5% period) | ± 0.25 ppm | | DC Current | | | 88.8 µA to 111.2 mA | ± (0.25% + 0.5 µA) | | | | | -88.8 µA to -111.2 mA | ± (0.25% + 0.5 µA) | | Square Current | | | 88.8 µApp to 111.2 mApp | ± (0.25% + 0.5 µA) | | Resistance | | | 10 Ohm to <40 Ohm | ± 0.5% | | | | | 40 Ohm to 90 Ohm | ± 0.1% | | | | | >90 Ohm to 150 Ohm | ± 0.5% | | | | | 50 kOhm to <800 kOhm | ± 0.5% | | | | | 800 kOhm to 1.2 MOhm | ± 0.1% | | | | | >1.2 MOhm to 12 MOhm | ± 0.5% | | Capacitance | | | >1 pF to 35 pF | ± (2% + 0.25 pF) | | | | | >35 pF to 95 pF | ± (3% + 0.25 pF) |
Composite Video Output: Amplitude: 1.0 Vpk-pk Pattern: (Full Raster) White, Grey or Black Sync Polarity: Positive or Negative Standards: 625-line 50 Hz, 525-line 60 Hz Trigger Output: Composite Sync or Odd Field start.
LF Linear Ramp: Waveforms: 1 Vpk-pk triangle, symmetrical about ground Ramp Time: 1 ms, 10 ms, 100 ms or 1 s ± 0.25 ppm (Ramp Time waveform is rising/falling/flat s over period 3 x s For example: 1 ms rising, 1 ms falling, 1 ms flat, over 3 ms period).
Overload Pulse: Amplitude: 5 V to 20 V into 50 Ohm Polarity: Positive or Negative from ground Duration: 0.2 s to 100 s (subject to pulse energy limits) Energy: 1.6 J to 50 J Power into 50 Ohm: 0.5 W to 8 W Trigger: Manual triggering only, maximum repetition rate 0.3 Hz.
Zero Skew: Unadjusted Skew: <±50 ps ch to ch Skew Temp Coefficient: <0.2 ps/°C Output Amplitude: 1 Vpp Frequency Range: 10 Hz to 100 MHz
Input Leakage Function: Open Circuit Output Leakage <±50 pA Short Circuit Output Offset <±15 µV
TOP (q) Voltage Simulation of Temperature
(1) Simulated Thermocouple Output and Thermocouple Measurement
| B | 600°C to <800°C | ± 0.44°C | | | 800°C to <1000°C | ± 0.34°C | | | 1000°C to <1550°C | ± 0.30°C | | | 1550°C to 1820°C | ± 0.33°C | | C | 0°C to <150°C | ± 0.30°C | | | 150°C to <650°C | ± 0.26°C | | | 650°C to <1000°C | ± 0.31°C | | | 1000°C to <1800°C | ± 0.50°C | | | 1800°C to 2316°C | ± 0.84°C | | E | -250°C to <-100°C | ± 0.50°C | | | -100°C to <-25°C | ± 0.16°C | | | -25°C to <350°C | ± 0.14°C | | | 350°C to <650°C | ± 0.16°C | | | 650°C to 1000°C | ± 0.21°C | | J | -210°C to <-100°C | ± 0.27°C | | | -100°C to <-30°C | ± 0.16°C | | | -30°C to <150°C | ± 0.14°C | | | 150°C to <760°C | ± 0.17°C | | | 760°C to 1200°C | ± 0.23°C | | K | -200°C to <-100°C | ± 0.33°C | | | -100°C to <-25°C | ± 0.18°C | | | -25°C to <120°C | ± 0.16°C | | | 120°C to <1000°C | ± 0.26°C | | | 1000°C to 1372°C | ± 0.40°C | | N | -200°C to <-100°C | ± 0.40°C | | | -100°C to <-25°C | ± 0.22°C | | | -25°C to <120°C | ± 0.19°C | | | 120°C to <410°C | ± 0.18°C | | | 410°C to 1300°C | ± 0.27°C | | R | 0°C to <250°C | ± 0.57°C | | | 250°C to <400°C | ± 0.35°C | | | 400°C to <1000°C | ± 0.33°C | | | 1000°C to 1767°C | ± 0.40°C | | S | 0°C to <250°C | ± 0.47°C | | | 250°C to <1000°C | ± 0.36°C | | | 1000°C to <1400°C | ± 0.37°C | | | 1400°C to 1767°C | ± 0.46°C | | T | -250°C to <-150°C | ± 0.63°C | | | -150°C to <0°C | ± 0.24°C | | | 0°C to <120°C | ± 0.16°C | | | 120°C to 400°C | ± 0.14°C |
Temperature standard is selectable between ITS-90 or IPTS-68. (2) Simulated RTD Output
| Pt 385, 100 Ohm | -200°C to <-80°C | ± 0.05°C | | | -80°C to <0°C | ± 0.05°C | | | 0°C to <100°C | ± 0.07°C | | | 100°C to <300°C | ± 0.09°C | | | 300°C to <400°C | ± 0.10°C | | | 400°C to <630°C | ± 0.12°C | | | 630°C to 800°C | ± 0.14°C | | Pt 3926, 100 Ohm | -200°C to <-80°C | ± 0.05°C | | | -80°C to <0°C | ± 0.05°C | | | 0°C to <100°C | ± 0.07°C | | | 100°C to <300°C | ± 0.09°C | | | 300°C to <400°C | ± 0.10°C | | | 400°C to 630°C | ± 0.12°C | | PtNi 385, 120 Ohm (NI120) | -80°C to <0°C | ± 0.08°C | | | 0°C to <100°C | ± 0.08°C | | | 100°C to 260°C | ± 0.09°C |
Temperature standard is selectable between ITS-90 or IPTS-68.
TOP (q) AC Harmonics
| 10 Hz to 45 Hz | 33 mV to 32.9999 V | 3.3 mA to 2.19999 A | 10 mV to 3.3 V | | 45 Hz to 65 Hz | 33 mV to 1000 V | 3.3 mA to 11 A | 10 mV to 3.3 V | | 65 Hz to 500 Hz | 33 mV to 1000 V | 33 mA to 11 A | 100 mV to 3.3 V | | 500 Hz to 1 kHz | 330 mV to 1000 V | 33 mA to 11 A | 100 mV to 3.3 V | | 1 kHz to 5 kHz | 3.3 V to 1000 V | 33 mA to 2.19999 A | 100 mV to 3.3 V |
Maximum frequency of harmonic output: 10 kHz. Harmonics available: 2 through 50, limited by maximum frequency of harmonic output. (For example: a fundamental frequency of 5 kHz limits the maximum harmonic to 2, i.e. 10 kHz. Harmonics through to 50 are available for a fundamental frequency of up to 200 Hz). BMC for amplitude (voltage or current) is as per the BMCs given in sections (a) DC voltmeters, (b) AC voltmeters, (c) DC ammeters, and (d) AC ammeters, except that the floor component is doubled. (Note that BMCs for Auxiliary Voltage are not given). For example: the BMC for the amplitude of a 230 V output at 50 Hz is: ± (0.05% + (6600 µV x 2)), which is ± 0.1282 V. BMC for phase is the greater of that given in (g) Phase angle indicators or 1°. For example: the BMC for phase for a 400Hz fundamental frequency and 10 kHz harmonic frequency is ± 10° (from Phase angle BMC for 10 kHz). Another example: the BMC for phase for a 50 Hz fundamental frequency and 400 Hz harmonic frequency is ± 1° (Phase angle BMC for 400 Hz is ± 0.9°, therefore ± 1° is used).
TOP (q) Electrical Test EquipmentThe BMCs in this section are applicable to the use of the Transmille 2100 for the calibration of the following equipment: Insulation testers, PAT testers, RCD testers, Line/Loop testers, Earth testers, and equipment employing one or more such functions. (1) Continuity - Resistance
| 20m Ohm | ± (0.25% + 0.01 Ohm) | | 0.1 Ohm to 20 Ohm | ± (0.25% + 0.01 Ohm) | | 100 Ohm | ± (0.25% + 0.01 Ohm) | | 1k Ohm | ± (0.25%+ 0.01 Ohm) |
(2) Continuity - Current
| 0 mA to 320 mA | ± (1.3% + 0.0006 Ohm) |
(3) Insulation Resistance - Resistance
0 MOhm to 5 MOhm | ± 0.1% | | >5 MOhm to <2 GOhm | ± 1% | | 2 GOhm to 10 GOhm | ± 5% |
(4) Insulation Resistance - Voltage
| 0 kV to 1.1 kV | ± (1% + 0.8 V) |
(5) Insulation Resistance - Current
| 0 mA to 9.999 mA | ± (1.3% + 6 µA) |
(6) AC Voltage Output
| 0 V to 400 V | ± (2% + 0.1 V) |
(7) RCD Time
(8) RCD Current
| 3 mA to <200 mA | <190 ms | ± (5% + 0.2 mA) | | | 190 ms to 5 s | ± (1.2% + 0.06 mA) | | 200 mA to 3 A | <190 ms | ± (5% + 2 mA) | | | 190 ms to 5 s | ± (1.2% + 0.6 mA) |
(9) Loop Resistance
| RANGES 1-9 (0.00 Ohm - 10 Ohm) | ± (0.5% + 0.022 Ohm) | | RANGES 10-11 (100 Ohm - 1 kOhm) | ± (0.5% + 0.058 Ohm) |
Resistance values are nominal. The Auto Loop BMC is ± 18 mOhm. This has been added to the Loop Resistance BMCs above. (10) PAT Earth Bond Resistance | RANGES 1-11 (0.00 Ohm - 1 kOhm) | ± (0.5% + 0.004 Ohm) |
Resistance values are nominal. (11) PAT Earth Bond Current
| 100 mA | ± (1.5% + 6 mA) | | 10 A | ± (1.5% + 0.06 A) | | 30 A | ± (1.5% + 0.06 A) |
(12) PAT Insulation Resistance | 0 MOhm to 5 MOhm | ± 0.1% | | >5 MOhm to <2 GOhm | ± 1% | | 2 GOhm to 10 GOhm | ± 5% |
(12) PAT Leakage Current
| 2 mA | ± (1.5% + 2 µA) | | 4.7 mA | ± (1.5% + 2 µA) | | 7.7 mA | ± (1.5% + 2 µA) |
(13) PAT Leakage Voltage
| 100 V to 300 V | ± (1.5% + 0.9 V) |
(14) PAT Load Testing
| 0 Ohm (Short Circuit) | ± 0.02 Ohm | | 440 Ohm (0.13 kVA) | ± 5% |
Note that 0.13 kVA load assumes 240 V supply, and as is given in manufacturer's specifications. Actual loading using 230 V supply is 0.12 kVA. (15) Line Voltage Measurement
| 200 V to 260 V | ± (0.8% + 0.6 V) |
TOP 5.91 Frequency Measurement and Time MeasurmentThese BMCs are applicable to the use of the Wavetek 9500. (a) Frequency Meters
| 0.1 Hz to <12 kHz | ± 3ppm (levelled sine voltage) | | 12 kHz to 1100 MHz | ± 0.25ppm (levelled sine voltage) | | 10 Hz to 100 kHz | ± 0.25ppm (square voltage) |
| 450.5 ps to 55 s | ± 0.25ppm (time markers) |
(c) Counters
| 0.02 Hz to 1100 MHz | ± 0.25ppm |
TOP |